The prototype provides mandatory breathing and mandatory breathing with patient triggering of early breaths with the following parameters and alarms. The prototype can accept input parameters requiring up to 70L/min of airflow. It will reject settings that exceed this amount.
Settable Parameters
Parameter | Value | Unit |
---|---|---|
Tidal Volume | 100-800 | mL |
Tidal Volume | 100-800 | mL |
Respiratory Rate | 10-40 | Breaths/minute |
I:E Ratio | 2:1 to 1:4 | Ratio |
Maximum Pressure | 100-600 | mm H2O |
Patient Triggered Breathing | On/Off | |
Alarm Pressure | 100-700 | mm H2O |
Alarm Low Minute Volume | 0-30000 | mL |
Alarm High Minute Volume | 0-30000 | mL |
Alarm Disconnect PIP | 0-200 | mm H2O |
Observable Parameters
In addition to the settable parameters, the following parameters can be observed on the LCD.
Parameter | Unit | Description |
---|---|---|
Tidal Volume | mL | Delivered on last breath |
Minute Volume | mL | Volume delivered over last minute |
PIP | mm H2O | Peak pressure on last breath |
PEEP | mm H2O | End expiratory pressure on last breath |
Alarms
Parameter | Description |
---|---|
Disconnect | PIP below threshold for 3 breaths |
Pressure | Alarm pressure threshold exceeded |
Low Minute Volume | Minute volume below set threshold |
High Minute Volume | Minute volume exceeds set threshold |